Humidity Sensing in Stability Chambers: Two Important Guidances, Redundant Humidity Sensors, and Intervals for Requalification

Laboratory test tubes
Piritta Maunu
Life Science Regulatory Compliance and Industry Expert​
Published:
Life Science
Stability testing chamber applications must meet the following criteria:
  • Proper documentation, including SOPs and periodical reports
  • Chambers and rooms are equipped with multiple sensors spread evenly throughout the controlled area
  • Generous multilevel shelving providing orderly storage and proper exposure to the controlled environment
  • Acceptable monitoring equipment (probes, data recorders, etc.)
  • Continuous recording of data and full traceability
  • Corrective action taken when stability factors go outside the specifications
  • Additionally, stability testing requires an alarm function that can detect and announce when an excursion from conditions defined in the study occurs. Pharmaceutical companies have adopted various methods for capturing and announcing abnormal conditions.
These include:
  • Alarms if monitored values go outside a predefined value
  • Alarms on excursion conditions being breached (usually a set temperature or humidity for a particular time)
  • SMS or e-mail alerts triggered by alarms or events
  • The FDA, CDER, CBER and the ICH collaborated to create "Guidance for Industry: Q1A(R2) Stability Testing of New Drug Substances and Products," first published in 1994 and revised in 2001, and 2003. This guidance defines what stability data package for a new drug substance or drug product is sufficient for a registration application within the three regions of the European Union (EU), Japan, and the United States. The guidance was adopted by Health Canada in 2003.

BLOG-IMAGE-LSH-CALIBRATION-LAB-VAISALA

 

Another source of guidance on stability testing is the World Health Organization, which has published
"Stability testing of active pharmaceutical ingredients and finished pharmaceutical products Annex 2" as part of their
Technical Report Series.

In combination, both of these documents offer the current guiding principles for stability testing protocols. If there are key takeaways from ICH and WHO guidance on testing humidity in stability chambers, it's that the sensor must suit the application; be positioned appropriately, and calibrated routinely over the range of humidity that will be studied. In addition to that, EU regulations require the following;

"The equipment used for the ongoing stability programme should be qualified and maintained following the general rules of EU GMP Volume 4, Chapter 3 (Premise and Equipment) and Annex 15 (Qualification and Validation)".

Read more about Humidity Sensors in Stability Testing by clicking the link below.

Stability Chambers & Redundant Monitoring Sensors

Like other sensors, humidity sensors may drift over time. As drift occurs, control loops continue to respond to sensor inputs, so it's possible that everything may appear to be in control and within specification. Adding one or more redundant sensors reduces the risk of undetected sensor drift. It's best to detect out of specification conditions before stability studies are compromised. The solution is often to install redundant sensors to offset risks to product safety and efficacy.
 

Independent monitoring sensors need to be highly accurate, stable, and (ideally) equipped to send alerts outside certain thresholds.  Most importantly, redundant sensors must be suitable for the intended conditions (e.g. 40°C / 75% humidity) and calibrated to ICH guidelines with a multi-point calibration.

Although sensor drift may occur too slowly to be detected, it can occur faster than most calibration cycles. Even small levels of contamination can cause significant drift. Calibrating once per year, or even every six months depending on how demanding your stability tests are, is expected.

How often do we need to ReQualify Stability Chambers?

Paul Daniel often receives questions about alarm limits and validation-mapping frequency for stability chambers. 

Paul writes:
In regards to setting alarm limits, it may not be as simple as it seems. The ICH guidelines are challenging at ±5%RH to the point that I have actually begun to think of the environmental uniformity of stability chambers as an uncertainty. In that light, one might use statistical techniques (like guard-banding) to set process limits like alerts. Here's a blog that touches upon that.

For re-validation intervals, I think it's critical to consider that even when operated under a good change control program, things wear out.  In the case of a stability chamber – motors wear, door seals warp, valves slowly clog, desiccant ages, and insulation hardens.  
 

It falls to a process experts to determine how often to qualify their stability chambers. It's a question of risk tolerance, and it's predicated on the quality guidelines of individual companies, their history with auditors, and most importantly; the sensitivity of their processes.

Personally, I think auditors are interested in knowing revalidation intervals for stability chambers. But, I also think that if you have data that indicates the systems you use for stability testing are very stable, auditors would accept a longer frequency of remapping. 

BLOG-IMAGE-SENSOR-VAISALA

ICH

ICH Quality Guidelines

WHO

Stability testing of active pharmaceutical ingredients and finished pharmaceutical products.

Humidity Sensors in Stability Testing

Vaisala stability chamber monitoring solutions provide independent real-time trends and audit trails with flexible alarming and customizable reporting. Vaisala reduces the risk of non-compliance with the requirements of the ICH and FDA for monitoring temperature, relative humidity, and other critical parameters.

 

 

Paul Daniel, Vaisala

Senior GxP Regulatory Compliance Expert

Paul Daniel has worked in the GMP-regulated industries for over 25 years helping manufacturers apply good manufacturing practices in a wide range of qualification projects.  His specialties include mapping, monitoring, and computerized systems.  At Vaisala, Paul oversees and guides the validation program for the Vaisala viewLinc environmental monitoring system.  He serves as a customer advocate to ensure the viewLinc environmental monitoring system matches the demanding requirements of life science and regulated applications.  Paul is a graduate of University of California, Berkeley, with a bachelor’s degree in biology.

Add new comment