Instrument Tolerances: Manufacturer vs. Process

Calibration
Profile picture for user janice.bennett
Janice Bennett-Livingston
Aug 24th 2013
Industrial Measurements
Life Science
Science & Sensing Technologies

This week's blog comes from Vaisala's Calibration and Repair Service department:

Using the instrument maker’s tolerance, there is often a higher risk of an “Out of Tolerance” appearing on a calibration certificate; this costs money. Consider the following scenario: You have just received back from calibration the set of temperature sensors used to monitor the warehouse. Most of the instruments have been shown to be within manufacturer published tolerances. A few of the instruments, though, are listed as out of tolerance. Now the fun starts! 

BLOG-IMAGE-Calibrated-parameters-Vaisala

The out-of-tolerance results need to be fully documented, including a full investigation into the use history of the instruments since their last calibration. All of this is rightfully required by your Quality Management system to ensure that no product has been negatively affected by the larger than expected errors in the readings. The investigation shows that the instrument manufacturer’s tolerances for the instrument are tighter than the tolerance required to monitor the warehouse. All of the out-of- tolerance points from the calibration certificates show that the readings of the instruments are within the tolerances for the warehouse monitoring system; therefore, no products have been negatively affected. 

This is a common situation that many of us find ourselves dealing with, a difference between manufacturer tolerances and process tolerances. In many cases, we have selected our process monitoring instruments based on more than just the measurement tolerances of an individual instrument.

We may have chosen an instrument with higher accuracy for a number of reasons, including: 

  • Compatibility with existing monitoring systems
  • Lower risk of an out-of-tolerance reading affecting products
  • Better overall value
  • The manufacturer was on the Approved Vendor


Whatever the reason, the manufacturer-specified tolerances are often significantly tighter than our process requires. This leaves us in a situation where we have a higher risk of an “Out of Tolerance” appearing on a calibration certificate than if the instrument being used had a tolerance matching our process tolerance. This in practice is costing extra money either due to an increase in calibration costs, a shortening of calibration intervals, or an increase in investigations of out-of-tolerance conditions.

There is a potential solution which will help contain costs and will also not increase the risk of negatively affecting product quality. When sending instruments out for calibration, have the calibration laboratory use the process tolerance when evaluating for in- or out-of-tolerance. The process tolerance is typically the more relevant limit rather than the manufacturer specification.

The process tolerance was established not by looking at what instruments are capable of performing, but by looking at the requirements of the process. It is this limit that tells us when the product may be affected and is, therefore, a more relevant tolerance to ensure that the measurement instrument is meeting.

Most calibration laboratories default to using manufacturer specifications when performing this evaluation. The calibration laboratory does not know what the process limits are for an instrument unless we tell them; the only information that is readily available at the calibration laboratory is the manufacturer specification. When selecting a calibration vendor, ask if they have the capability to use customer specified acceptance limits.



In most cases, this should not pose a significant problem. When sending the instruments in for calibration, including the instructions for the values to use for the process limits. Using process limits instead of manufacturer specifications provides a low-risk and low-cost mechanism for reducing the cost and effort associated with out-of-tolerance events. It provides a more relevant analysis of the calibration results and will limit the need to reduce calibration intervals in the case of equipment not meeting manufacturer specifications.

 

Learn More About Vaisala Calibration Services

Add new comment

Calibration eBook

This book serves as a generic introduction to calibration. We discuss the rational behind calibration, what are the factors that affect the need to calibrate. The book also contains specific information on calibration of relative humidity, dewpoint tempera...

Vaisala´s Calibration Services

Standard Calibration

Standard Calibration

Standard calibration performed on Vaisala instruments before they leave the factory. In order to maintain best accuracy for the measurement, the...
Accredited Calibration

Accredited Calibration

Many industries are required by regulatory bodies to comply with strict quality standards. Vaisala offers calibration services accredited according...
Custom Calibration

Custom Calibration

Custom calibration is the correct choice if specific points within your instrument’s measurement range are needed. Calibrating and adjusting your...
On-Site Calibration

On-Site Calibration

Calibrating Vaisala Continuous Monitoring Systems (CMS) on site helps to maximize reliability and profitability by minimizing costly equipment...
Vaisala Service Center Repair

Service Center Repair

Our professional repair services help extend the life of your Vaisala equipment. Repairs include fault tracing and replacement of failed components...