Vaisala BAROCAP® Barometer PTB110

  • Vaisala BAROCAP® sensor
  • Several pressure ranges
  • Accuracy ±0.3 hPa at +20 °C
  • Long-term stability
  • On/off control with external trigger
  • Output voltage 0 ... 2.5 or 0 ... 5 VDC
  • Current consumption less than 4 mA
  • Mountable on a (35 mm wide) DIN rail
  • NIST traceable (certificate included)
 

Operating range (1 hPa = 1 mbar)

 

Pressure ranges
    




Total accuracy at +20 °C

Sensor

500 … 1100 hPa
600 … 1100 hPa
800 … 1100 hPa
800 … 1060 hPa
600 … 1060 hPa

± 0.3 hPa

BAROCAP®

Operating environment

 

Temperature range
Humidity range

-40 ... +60 °C (-40 ... +140 °F)
non-condensing

Inputs and outputs

 

Operating voltage
Power consumption:
     operation mode
     shutdown mode
Output voltage

10 ... 30 VDC

less than 4 mA
less than 1 µA
0 ... 2.5 VDC
0 ... 5 VDC

Mechanics

 

Weight
Material (housing)

90 g
plastic cover, ABS/PC blend

For complete specifications, please see the brochure or user guide in "Product Documentation".

 

 

Typical delivery times

 

  USA

  EUROPE

  JAPAN, CHINA, ASIA

  1 - 2 products

  4 work days *

  3 work days *

  4 work days *

  1 - 5 products

  7 work days

  7 work days

  7 work days

  6 or more products

  21 work days

  21 work days

  21 work days


Note that the products are configured to your needs.

* Urgent delivery fee will be added.

 

Calibration services

 

Calibrated quantities
Traceability

barometric pressure
NIST

 

Calibration services are available at Vaisala service centers for a fixed price.
-Accredited calibrations (ISO17025)
-Factory calibrations (ISO9001)

Service price includes:
-calibration and certificate before adjustment
-adjustment to meet specifications
-calibration and certificate after adjustment
-replacement of wearing parts

Spare parts

 

Common spare and wearing parts are available. For availability contact sales.

Repair

 

Repair services are available at Vaisala service centers for a fixed price.

Repair price includes:
-repair work
-required parts
-calibration certificate
-service report

Technical support

 

Technical support for your installed units, contact helpdesk@vaisala.com.

 

 

 

 