Performance |
| Measurement range
Accuracy (including noise,
linearity, and repeatability) Temperature dependence
over T range
Stability
Zero drift
Response time of
measurement
Diffusion limited response in
still air without filters with stainless steel mesh filter with stainless steel mesh and PTFE filters
| 0 ... 100 % O2
±0.2 % O2
±2 % of reading
±1 % of reading/year
±0.1 % O2/year
3 s
T63/T90 10 s / 20 s
10 s / 25 s
30 s / 70 s
|
Operating environment |
| Operating temperature range for probe (in-line
installation) for electronics (housing) for transmitter (ambient gas measurement)
Operating pressure range
Maximum pressure range for
probe
|
-20 ... +80 °C
-40 ... +60 °C
-20 ... +60 °C
0.8 ... 1.4 bar
up to 10 bar
|
Inputs and outputs |
| Power supply input range
Power consumption maximum typical
Analog output maximum load
Serial output (2-wire non-
isolated)
Alarm/control relay |
11 … 36 VDC
6 W
3 W
0/4 … 20 mA, sourcing
500 Ω
RS-485
30 VAC/60 VDC
|
Mechanics |
| Housing material
Probe material
Housing classification
Weight
Mounting flance diameter Can
be fitted to standard flanges. Minimum sizes: DIN (2527B) DN50 and (150) 2.5" Cable bushing
Stainless steel mesh filter
Wetted materials
AISI 316, EPDM or Kalrez®
(optional), PTFE (optional), SiN, MgF2 | G-AlSi10 Mg (DIN
1725)
AISI 316
IP66
2.2 kg
96 mm
Cable gland M20 x 1.5
openings Ø 310 μm |
For complete specifications, please see the brochure