Performance |
|
|
Measurement range Accuracy (including noise, linearity, and repeatability) Temperature range over T range Stability Zero drift Response time of measurement Diffusion limited response in still air without filters with stainless steel mesh filter with stainless steel mesh and PTFE filters Pressure dependence without pressure compensation 0.8 ... 1.2 bar 1.2 ... 1.4 bar Accuracy of pressure compensation Background gas effects for CO2 and H2O uncompensated
Accuracy of background gas compensation 0 ... 50 vol-% CO2 0 ... 300 g/m3 H2O (Td=80°C ) |
0 ... 100 % O2
±0.2 % O2
±2 % of reading ±1 % of reading/year ±0.1 % O2/year
3 s
T63/T90 10 s / 20 s
10 s / 25 s
30 s / 70 s
-2% of reading -5% of reading
±0.25% of reading
<1% of reading for <6 vol -% CO2 <1% of reading for gas dew point < 30 °C
±-0.5 % of reading ±1% of reading |
Operating environment |
|
|
Operating temperature range for probe (in-line installation) electronics (housing) (Kalrez® seals) transmitter (ambient gas measurement) Storage temperature range Operating pressure range Maximum pressure range for probe |
-20 ... +80 °C -20 ... +60 °C
-20 ... +60 °C -55 …+80°C 0.8 ... 1.4 bar up to 10 bar
|
|
|
Compliance IEC(EN)-61326 Electrical equipment for measurement, control and laboratory use. EMC requirements.
|
|
Laser safety OMT364 classifications for the transmitter USA and Canada
OMT364 classifications for the probe USA and Canada |
Class 1 laser product
FM, Non-incendive for Class I, Division 2, Groups A, B, C, and D. Ta=-20°C to +60°C
FM, Suitable for Class I, Division 1, Groups A, B, C, and D. Ta=-20°C to +80°C
|
Inputs and outputs |
|
|
Power supply input range Power consumption maximum typical Analog output maximum load Serial output (2-wire non- isolated) Alarm/control relay |
11 … 36 VDC
6 W 3 W 0/4 … 20 mA, sourcing 500 Ω
RS-485 30 VAC/60 VDC
|
Mechanics |
|
|
Housing material Probe material Housing classification Weight, transmitter and probe Weight, sample cell Mounting flance diameter Cable bushing Stainless steel mesh filter Wetted materials AISI 316, Kalrez® PTFE (optional), SiN, MgF2 |
G-AlSi10 Mg (DIN 1725) AISI 316 IP66 2.2 kg 2.6 kg 97 mm Cable gland M20 x 1.5 openings Ø 310 μm |