Performance |
| CO2 measurement range
Accuracy (including repeatability, non-linearity and calibration uncertainty)
Long-term stability
Response time T90 Temperature
dependence, typical
Pressure dependence, typical
Warm-up time
Product lifetime | 0 … 2000 ppm
± (2.5 % of range + 3 % of reading) ± 100 ppm CO2 / 5 years
1 min
-0.35 % of reading / °C
+0.15 % of reading/ hPa
1 min, 10 min for full specification
> 10 years |
Operating environment |
| Temperature
Humidity
Pressure
Electromagnetic compatibility | -5 ... +45 °C (23 …
113 °F)
0 ... 85 % RH
700 … 1200 hPa
EN61326-1. Generic Environment |
Inputs and outputs |
| Operating voltage
Power consumption
Outputs | 24 V (±20 %) AC/DC
<2 W
4 … 20 mA, 0 … 10 V, RS-485, 2-wire, non-isolated |
For complete specifications, please see the brochure
or manual in "Product Documentation".