What is the response time of the moisture in oil sensor?

The thin-film polymer capacitive sensor is extremely responsive to a small step change in moisture level. This can be demonstrated by grasping the probe in your hand. The moisture from your skin will be detected within 2-3 seconds. However, this response time is diminished in oil due to the oil’s viscosity and the slow rate of travel of water through the oil. In still oil at 20 °C, 90% of the reading will be reached in ten minutes.  This response time dramatically improves by installing the probe in a flowing oil line. Since the sensor detects water by absorption, the greater the flow rate, the faster the response time.


This page was printed from www.vaisala.com on May 13, 2008
URL: http://www.vaisala.com/instruments/knowledgecenter/faq/faqs/what%20is%20the%20response%20time%20of%20the%20moisture%20in%20oil%20sensor.html