What is the purpose of the chemical purge option on the HUMICAP
sensor?
In some applications the sensor gain may decrease gradually due to an
interference caused by a particular chemical present in the measured gas. The
sensor polymer absorbs the interfering chemical and this reduces the ability of
the polymer to absorb water molecules; consequently the sensor gain decreases.
In chemical purge the humidity sensor is heated to a temperature of
approximately +160 °C for several minutes, which evaporates the interfering
chemical. The chemical purge cycle takes approximately 6 minutes.
The decrease of sensor gain due to an interfering chemical and the effect of
the chemical purge process are illustrated in the graph below.