How will dust and dirt affect the measurement for the SPECTRACAP® oxygen
sensor?
The SPECTRACAP® sensor uses an optical signal to measure oxygen and requires
adequate optical transmission through two different surfaces that are exposed
to the gas being measured. There are two types of protection against the
effects of dust and dirt in the SPECTRACAP® sensor. The first is mechanical
filtering. The probe is equipped with a stainless steel mesh filter (300 µm
filtering) and an optional hydrophobic PTFE filter (average pore size 8 µm).
The second type of protection is the measurement algorithm itself, which can
tolerate up to 90 % loss of optical transmission without interrupting the
measurement. The OMT355 will issue a warning signal when optical transmission is
reduced to 50%.