Maximum characteristics |
|
Input voltage (across channel line pair and from line to GND, terminals #3) |
|
|
power channels |
max. ± 43 V |
|
data channels |
max. ± 13 V |
|
Input common mode voltage any line to earth |
max. ± 72 V |
|
Throughput current |
|
|
power lines |
max. 1.5 A |
|
data lines |
max. 0.16 A |
|
Throughput resistance (per line) |
|
|
power lines |
0.3 ohms |
|
data lines |
15 ohms |
|
Turn-on voltage |
|
|
power channels |
max. ± 60 V |
|
data channels |
max. ± 16 V |
|
Surge current |
|
|
to earth |
max. 10 kA |
|
differential |
max. 5 kA |
|
EMC surge tolerance |
EN 61000-4.5, (4 kV, 2 kA) IEEE C62.45 (6 kV, 3kA) |
General |
|
Operating and storage temperature |
-52 ... +70 °C (-60 ... +158 °F) |
|
Installation and maintenance work temperature |
-40 ... +70 °C (-40 ... +158 °F) |
|
Dimensions |
|
|
housing (w x h x d) |
|
|
with cable glands and |
130 x 94 x 58 mm |
|
mounting assembly |
130 x 120 x 69 mm |
|
cables |
Ø 4 ... 8 mm |
|
wires |
Ø 0.4 ... 1.7 mm (AWG 26 … 14) |
|
Housing material |
polycarbonate, stainless steel |
|
Weight |
0.65 kg |
For complete specifications, please see the brochure or manual in "Documents".