Vaisala HUMICAP® Humidity and Temperature Transmitter Series HMW90

HMW90 HMW90D
 

 Buy Online

 
 

 Contact Us

 


 

Learn More

CEN_rightcolumncode2
    Overview
     

     Overview

     
    • Easy installation, configuration,and field adjustment
    • Vaisala HUMICAP® 180RSensor with superior long-term stability
    • Humidity parameter options: relative humidity, dew point, and mixing ratio
    • Full 0 … 100 %RH measurement range
    • Up to ±1.7 %RH accuracy
    • Special output scaling available
    • One relay (HMW93 model only)
    • UL V0 materials
    • User exhangeable humidity and temperature module
    • NIST traceable calibration
    Technical Specifications
     

     Technical Specifications

     

    Relative humidity

     

    Measurement range
    Accuracy 
         Temperature Range
          0...90 %RH
          90...100 %RH
          Temperature range

          0 ... 90 %RH
          90 ... 100 %RH
    Stability in typical
    HVAC applications
    Humidity sensor 

    0 ... 100 % RH

    +10 ... 40 °C (+50 ... +104 °F)
    ±1.7%RH
    ±2.5 %RH
    -5 ... +10 °C, +40 ... + 55 °C
    (+23 ... +50 °F, +104 ... +131°F)
    ±3 %RH
    ±4 %RH

    ±0.5 %RH/year
    Vaisala HUMICAP® 180R

    Temperature

     

    Measurement range
    Accuracy
         +20 ... +30 °C
          (+68 ... +86 °F)
         +10 ... +20 °C, +30 ... +40°C
         (+50 ... +68 °F, +86 ... +104 °F)
         -5 ... +10 °C, +40...+55°C
         (+23 ... +50 °F, +104 ... +131 °F)
    Temperature sensor

    -5 ... +55 °C (+23 ... +131 ºF)


    ±0.2 °C (± 0.36 °F)

    ±0.3 °C (± 0.54 °F)

      ±0.5 °C (± 0.90 °F)
    Digital temperature sensor

    Operating environment

     

    Operating temperature range
    Storage temperature range
    Electromagnetic compliance
        

    -5 ... +55 °C (+23 ... +131 °F)
    -20 ... +60 °C (-4 ... +140 °F)
    Complies with EMC
    standard EN61326-1, Industrial Environment

    Inputs and outputs

     

    HMW92 and HMW92D
    Outputs
    Loop resistance
    Supply voltage

    Isolation between output channels
    HMW93, HMW93D
    Outputs
    Load resistance
    Supply voltage

    Max. current consumption

    Relay
    Service port


    2 x 4 ... 20 mA, loop powered
    0 ... 600 Ω
    20 ... 28VDC at 500 Ω load
    10 ... 28VDC at 0 Ω load
     500VDC

    2 x 0 ... 5V, 0 ... 10 V
    10 kΩ min
    18 ... 35 VDC
    24 VAC ±20 % 50/60 Hz
    12 mA
    max. with relay 25 mA
    1 pcs (max. 50 VDC, 500 mA)
    RS-485 line for temporary service use

    Mechanics

     

    IP class
    Housing colour
    Housing material

    IP30
    RAL9003
    ABS/PC
    UL-V0 approved

    For complete specifications, please see the brochure or user guide in "Documents".

     