Performance |
|
Measurement ranges GMT221
|
0 ... 2 % 0 ... 3 % 0 ... 5 % 0 ... 10 % 0 ... 20 %
|
|
GMT222 for low concentrations |
0 ... 2000 ppm 0 ... 3000 ppm 0 ... 5000 ppm 0 ... 7000 ppm 0
... 10 000 ppm
|
|
Accuracy (incl. repeatability, non-linearity and calibration
uncertainty) at +25 °C (+77 °F) and 1013hPa |
|
|
GMT221 |
± (1.5% of range + 2 % of reading) (applies for concentrations above
2% of full scale) |
|
GMT222 |
± (1.5% of range + 2 % of reading) |
|
Sensor |
Vaisala CARBOCAP® |
Operating environment |
|
Temperature |
-20 ... +60 °C (-4 ... +140 °F) |
|
display |
0 ... +50 °C (+32 ... +122 °F) |
|
Relative humidity |
0 ... 100 %RH non-condensing |
Inputs and outputs |
|
Outputs |
0 ... 20 or 4 ... 20 mA and 0 ... 10V
|
|
Operating voltage |
nominal 24 VAC/VDC |
|
Power consumption |
<4 W |
Mechanics |
|
Probe cable length |
2 m and 10 m (optional) |
|
Weight |
|
|
GMT221 |
max. 280 g |
|
GMT222 |
max. 300 g |
|
Material |
|
|
transmitter body |
ABS plastic |
|
probe |
PC plastic |
|
Housing classification |
IP65 (NEMA 4) |
Options and accessories |
|
GMP221, GMP222
|
spareprobe (use the order form to define
measurement range etc.) |
|
25245GM |
clips (2 pcs) for attaching the probe |
|
GM45156 |
mounting flange for the probe |
|
25665GMSP |
2.0 m probe cable |
|
210848GMSP |
10.0 m probe cable |
|
GMK220 |
calibration kit for interchangeable probes |
|
19040GM |
serial COM adapter for maintenance purposes
|
|
26150GM |
chamber for pump aspirated sampling systems |
|
GM70 |
hand-held CO2meter for field verification and calibration |