Carbon dioxide measurement |
|
Measurement range
Accuracy (including repeatability, non-linearity and calibration uncertainty) Sensor |
0 ... 2000 ppm (nominal; can be calibrated for other ranges: 0 ... 5000 ppm, 0 ... 10 000 ppm, 0 ... 20 000 ppm)
±(2 % of range + 2.0 % of reading) Vaisala CARBOCAP®
|
Temperature measurement (optional with GMW21) |
|
Measurement range Accuracy at +25 °C Sensor |
0 ... +50 °C (+32 ... +122 °F) ± 0.5 °C (± 0.9 °F) Semiconductor IC
|
Operating environment |
|
Temperature Humidity |
-5 ... +45 °C (+23 ... 113 °F) 0 ... 85 %RH non-condensing
|
Inputs and outputs |
|
Outputs Optional outputs Operating voltage Power consumption |
0 ... 20 mA or 4 ... 20 mA and 0 ... 10 V relay, LonWorks® interface nominal 24 VAC/DC (18 ... 30 VDC) <2.5 W
|
Mechanics |
|
Weight GMD20(D) GMW21 (D) GMW22 (D) Housing material Housing classification (electronics housing GMD20) |
140 g (170 g) 100 g (130 g) 90 g (120 g) ABS plastic
IP65 (NEMA 4)
|
Options and accessories |
|
GMI21 GMR20 GML20
GML20T
19222GM GMA20T
GM70 w/pump option |
Display and relay option for GMD/W21/22 Relay output option LonWorks® module with CO2 signal (not available when the display option is added) LonWorks® module with both CO2 and temperature signals (not available when the display option is added) Calibration software kit (disk serial COM adapter) Analog temperature module for GMW21 (not available when the display option is added) Hand-held CO2 meter for field verification |
For complete specifications, please see the brochure or user guide in "Documents".