Vaisala Veriteq SP 2000 Temperature Humidity Loggers

​Vaisala Veriteq SP-2000 precision temperature and humidity data loggers are compact, powerful and easy-to-use recorders for monitoring critical humidity-sensitive products and processes, such as those in labs, cleanrooms, and stability chambers.

Our self-powered humidity data loggers include on-board humidity & temperature sensors, non-volatile memory, 10-year battery and a precision clock.

Ideal for use in standalone or networked applications, the 2000 data recorder connects directly to a PC with USB or installs to your existing network via Ethernet, Power over Ethernet, or WiFi.

The SP-2000 also features an optional external channel with current or voltage inputs to record parameters such as differential pressure, C02, level, particles, or conductivity. An optional Boolean channel connects to door switches or alarm contacts.

SP-2000 data loggers are easy-to-use alternatives to chart recorders and bulkly data acquisition systems. Self-contained and battery-powered, these data loggers can be placed anywhere you need to measure temperature and humidity. No charts or wires are required.
 

 Contact Us

 


 

Learn More

LSH_rightcolumn_price_chat_adv_tech
    ​Features
     

     ​Features

     
    • ​Portable self-contained data loggers with internal temperature and humidity sensors, on-board memory and ten-year battery
    • Traceable instrument-grade RH and temperature measurement accuracy with adjustable time-based digital recording and multi-year recording capacity
    • Can be used for standalone data logging, continuous monitoring, or validation/mapping
    • NIST-traceable, A2LA accredited calibration. ICH and custom calibration points available
    • Flexible connectivity options include WiFi, ethernet network, Palm PDA connectivity
    • Loggers contain powerful Vaisala RH-sensors.  Thermoset polymer-based capacitive humidity sensor offers outstanding resistance to air-borne contaminants and condensation, delivering exceptional in-calibration service and reliability
    • Accurate to +/- 1 percent RH and .10 degree C, and able to sense changes .05% RH and .02 degrees C
    • Optional external channel for voltage or current inputs allow recording of differential pressure, level, particles, conductivity or C02
    • Optional Boolean channels for door switches or alarm contacts
    • Adjustable logging intervals with multi-year data recording capacity
    • Interface with Vaisala Veriteq Spectrum software or Vaisala Veriteq viewLinc software, providing the ability to view, download, store, and graph condition information. Can also be used with the Vaisala Veriteq Excel add-in, allowing you to work with the data in MS-Excel
    • See also our Data Logger FAQ
    Technical Specifications
     

     Technical Specifications

     


    Internal Relative Humidity Sensor

    Type of sensor Integrated high-accuracy capacitive polymer RH sensor
    Initial Accuracy

    ±1%RH over 10%RH to 90%RH at +20ºC to +30ºC
    ±2%RH over 10%RH to 90%RH at -20ºC to +70ºC

    Operating range -35°C to +85°C, 0% RH to 100% RH (non-condensing)
    Resolution 0.05% RH
    1-year Accuracy

    (*denotes Accuracy One Year
    after original deployment)

    *±2% RH over 10% RH to 90% RH at +20°C to +30°C
     
    ±3% RH over 10% RH to 90% RH at -20°C to +70°C
    Calibration Interval Annual (recommended)


    Internal Temperature Sensor

    Initial Accuracy ±0.10ºC over +20ºC to +30ºC

    ±0.15ºC over -25ºC to +70ºC
    Operating range -35 to +85°C (-31 to 185°F)
    Resolution 0.02°C at +25°C
    1-year Accuracy ±0.15°C over 20°C to +30°C
    ±0.25°C over -25°C to +70°C


    General Humidity Logger Specifications

    Size 3.4 x 2.3 x 1" (85x59x26mm);
    76g (2.7 oz)
    Operating Range -35 to 85°C (-31 to 185°F)
    and 0% RH to 100% RH
    Interfaces RS-232 serial, USB, Ethernet, WiFi, PoE network interface available.
    Mounting Magnetic strips; 3M Dual Lock™ fasteners
    Software Compatible with the following software packages: Vaisala Veriteq Spectrum, Vaisala Veriteq viewLinc, Spectrum Excel Add-in, Spectrum API (for custom applications) and OPC Server to add loggers onto compatible monitoring systems.
    Clock Accuracy: ± 1 min./month at -25° C to 70°C.
    Electromagnetic Compatibility

    FCC Part 15 and CE
    EN55022:2006
    EN61000-4-2:2001
    EN 61000-4-3:2006

    Power Source Internal lithium battery with life of 10 years at 1 min. sampling rate or longer.

    Memory

    Memory Type Non-volatile EEROM
    Data Sample Capacity 122,197 12-bit samples
    Memory Modes User-selectable wrap (FIFO) or stop when memory is full. User-selectable start and stop times.
    Memory Protection Data retention > 20 years backup without power.
    Sampling Rates User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day. Note: 10-year battery life is specified at 1 minute sample rate. Shorter sample intervals will affect battery life.


    Channel Configuration & Recording Span

    Model Number
    Channel Types
    CH1
    CH2
    CH3
    CH4
    2000-3CR
    Temperature
    RH
    Current 4 to 20 mA
     
    2000-35R
    Temperature
    RH
    Voltage 0 to 5 VDC
     
    2000-3AR
    Temperature
    RH
    Voltage 0  to 10 VDC
     
    2000-4BR
    Temperature
    RH
    Boolean
    Boolean
     
     
     
     
    Sample Interval
    10 seconds
    14.1 days
    7.1 days
    4.7 days
    3.5 days
    1 Minute
    2.8 months
    1.4 months
    23.8 days
    21.2 days
    5 Minutes
    1.2 years
    7.1 months
    4.7 months
    3.5 months
    15 Minutes
    3.5 years
    1.7 months
    1.2 years
    10.6 months
    1 hour
    13.9 years
    7.0 years
    4.6 years
    3.5 years


    Current Loop & Voltage Inputs

    Input Type

    Current Loop

    Analog Voltage

    Available Ranges
    0 to 22mA
    0 to 5VDC, 0 to 10VDC
    Resolution
    5.5μA
    0.025%F.S.
    Accuracy
    ±0.15% at +25°C (+77°F)
    ±0.15% at +25°C (+77°F)
    Input Impedances
    >1MOhms
    Isolation
    One common for logger
    One common for logger
    Overload Protection
    40 mA max. (reverse-polarity protected)
    ±24 VDC max. (reverse-polarity protected)

    *Temperature channel must be enabled when the RH channel is enabled.

    Termination resistance plus approximately 0.4 volt drop through a protection diode.

    Related Services
    • Service Center Calibration 
    • Regular calibration ensures that your Vaisala equipment continues to provide accurate, high-quality data. We offer traceable, standards-compliant calibration services for almost all Vaisala products and partner with world-class couriers for rapid, trouble-free return of your equipment.
    • On-Site Calibration 
    • On-site calibration helps you maximize both reliability and profitability by minimizing costly equipment downtime and ensuring compliance with relevant regulations and standards. Our expert engineers can also replace a range of components, carry out minor repairs, and provide the latest upgrades.
    • Repair Services 
    • Vaisala repair services include fault tracing and repair or replacement of failed components, as well as final testing to ensure your equipment is functioning properly. Repairs can be carried out on site or at one of our service centers.
    • Spare Parts 
    • Get rapid access to the spare parts you need to maximize the uptime of your equipment and ensure reliable and accurate operation. In addition to standard spare part delivery, we also offer an express service where parts are shipped within one business day.
    • Warranty 
    • All Vaisala products and services are covered by a 12-month warranty as standard, with selected products covered for up to 24 months. During this period you benefit from free repair or replacement of faulty equipment. With Vaisala’s extended warranty, you get free repair or replacement for five years.
    Documents