Vaisala Veriteq Temperature Data Recorder Series SP 1200

Vaisala Veriteq SP-1200 low temperature data loggers are perfect for monitoring, alarming, and recording environmental conditions in non-GxP-regulated environments like refrigerated warehouses, cold rooms, freezers and other low temperature applications.

The 1200 series can take up to 2 channels within a measurement range of -55C to +40C (-67°F to +104°F). Channel options are one external probe and one internal thermistor, or two external probes.
 

 Contact Us

 


 

Learn More

LSH_rightcolumn_price_chat_adv_tech
    Features
     

    Features

     

    The SP-1200 loggers are used in conjunction with Vaisala Veriteq viewLinc monitoring software or vLog SP reporting software to provide a simple, secure solution for temperature recording in a variety of cold environments.  Each palm-sized logger has on-board power and memory for autonomous recording that is immune to any network or power failure.

    Vaisala Veriteq viewLinc Continuous Monitoring software
    allows you to monitor conditions for all SP-1200 data loggers via wired or wireless ethernet. PoE connectivity is also available. The Vaisala Veriteq vLog SP software allows you to graph, track, download and display logger data, simply and quickly. Reports are customizable to your application requriements. For validated applications, we recommend VL-1200 data loggers combined with vLog VL.

    Technical Specifications
     

    Technical Specifications

     


    Size                                                  
    85 x 59 x 26 mm (3.4 x 2.3 x 1") 76 g (2.7oz)

    Interfaces                                       
    RS-232 serial, USB, Ethernet, WiFi, Power over Ethernet.

    Mounting
    Magnetic strips, 3M Dual Lock™ Fasteners; Snap-in connector provide secure probe connections

    Software

    • vLog SP for SP-series
    • vLog VL for VL-series 
    • viewLinc CMS for SP/VL-series 
    • OPC server for data recorder interoperability with OPC-compatible monitoring systems 
    Internal Clock Accuracy              
    +/-1 min./month over -55 °C to +40 °C (-67 °F to +104°F)

    Electromagnetic Compatibility
    FCC Part 15 and CE

    Power Source
    Internal 10-year lithium battery
    (Battery life specified with sample interval of 1 min. or longer)

    Applications
    ▪ Ideal for in situ cold temperature validation to -55 °C
    ▪ Can record with up to 2 external probes calibrated to -55 °C
    ▪ Highly accurate for use in GxP low temperature applications:
    Accuracy to +/-0.25

    External
    Temperature Sensors
    • Sensor Type                                                        L Range Probes

    • Sensor Tips

                    Stainless Steel                                   Diameter 3.2 mm (1/8")
                                                                                   Length 38 mm (1.5")
                    Sealed Teflon Tip                              
    (Waterproof)
                                                                                   Diameter: 3mm (0.12")
                                                                                    Length 28 mm (1.1")

    • Cable construction 
      2 mm (0.07") Diameter, Teflon coated cable
       
    • Probe length
      7.6 m (25')
    • Connector Color Code
      Green

    Temperature Probe Accessories
    Thermal Dampening Block, for use in refrigerators and freezers, simulates a glycol bottle to reduce fluctuations generated by opening and closing a door.

    Measurement Accuracy
    Internal Sensor & External Probe

    • Calibrated Measuring Range
      -55 °C to +40 °C (-67 °F to +104 °F) 
    • Standard Calibration points
      ** -55°C, -40°C, -20°C, -0°C, +20°C, +40°C
      ( -67°F, -40°F, -4°F, +32°F, +68°F, 104°F) 
    • Resolution
      0.02 °C at -20 °C (0.04 °F at -4 °F)

    • Initial Accuracy*
      +/-0.25 °C over -55 °C to +40 °C  +/-0.45 °F over -67 °F to +104 °F

    • One year Accuracy*  
      +/-0.42 °C over -55 °C to +40 °C (+/-0.76 °F over -67 °F to 104 °F)

    • Operating/Storage Range 
      -60 °C to +85 °C (-76 °F to +185 °F)

    *The accuracy specification for external channels is for a probe calibrated to the specific channel of the data logger and with the data logger at -55 °C to +40 °C (-67 °F to +104 °F)

    **Customized calibration points are available, from -60°C to +45°C
    (-76°F to +113°F)


    Internal Temperature Sensor
    Type of Sensor: Precision-tolerance epoxy-encapsulated NTC thermistor

    Recording Span:
                                                          
    Number of Channels Enabled
    Sample Interval                           1                                          2
    10 Seconds                               5.5 Days                             2.7 Days
    1 Minute                                    1.1 Months                          16.7 Days
    5 Minutes                                 5.5 Months                           2.7 Months
    15 Minutes                               1.3 Years                             8.3 Months
    1 Hour                                       5.4 Years                             2.7 Years

    The 1200 data recorder comes in the following models:

    • SP- or VL-1200-21L - One internal temperature channel and one external temperature channel.

    • SP- or VL-1200-22L - Two external temperature channels.