General
| Size |
3.4 x 2.1 x 1" (85 x 59 x 26mm); 60g (2.7 oz) |
| Operating Range |
-95°C to +70°C (-139°F to 158°F) through probes |
| Interfaces |
RS-232 serial port; Ethernet; PoE & WiFi interfaces available |
| Mounting |
Magnetic strips; 3M Dual Lock™Fasteners; Snap-in connector locks provide secure probe connections |
| PC Software |
- viewLinc automated monitoring, alarming & reporting
- OPC server software to add Vaisala Veriteq loggers to any OPC-compatible monitoring system
|
| Clock |
Accuracy: ±1 min./month at 0°C to +50°C (+32°F to +122°F) |
| Electromagnetic Interference |
Meets FCC Part 15 for digital devices; meets CE requirements for radiated emissions, electrostatic discharge, and radiated susceptibility. |
| Power Source |
Internal lithium battery with life of 10 years at 1 minute sampling rate. |
Multi-Application Temperature Recorder Accuracy
| Sensor |
Calibrated Measuring Range |
Resolution |
One year Accuracy |
| V Range External Probe |
-90°C to +50°C
(-130°F to +122°F) |
0.01°C at +25°C
(0.02°F at +77°F) |
±0.35°C over -90°C to +50°C
(±0.63°F over -130°F to +122°F) |
Memory
| Memory Type |
Non-volatile EEROM |
| Data Sample Capacity |
1016 - 68,600 16-bit samples (2 channels)
1416 - 101,375 16-bit samples (4 channels) |
| Memory Modes |
User-selectable: Wrap when memory full (FIFO) or Stop when memory full. User-selectable start time and (for VL series) stop time. |
| Memory Protection |
Data retention > 20 years backup without power. |
| Sampling Rates |
User-selectable rates from once every 10 seconds to once a day. Note: 10 year battery life is specified at a 1 minute sample rate. |
| Recording Span |
Recording span depends on sample interval selected and number of channels enabled. Chart below details some typical sampling rates and length of time logger will retain data in memory before wrapping around or stopping (see Memory Modes). |
Recording Span: 1016 - 22V
1 minute |
1.5 Months |
23.8 Days |
5 minutes |
7.6 Months |
3.8 Months |
15 minutes |
1.9 Years |
11.5 Months |
1 hour |
7.8 Years |
3.9 Years |
Recording Span: 1416 - 44V
1 minute |
2.3 Months |
1.1 Months |
23.5 Days |
17.6 Days |
5 minutes |
11.3 Months |
5.6 Months |
3.7 Months |
2.8 Months |
15 minutes |
2.8 Years |
1.4 Years |
11.3 Months |
8.5 Months |
1 hour |
11.5 Years |
5.7 Years |
3.8 Years |
2.8 Years |
Temperature Probes: Stainless or Teflon Tip
V Range external probe:EPT-23N-XXV |
-95°C to +70°C
(-139°F to +158°F)
|
|
Stainless Steel Diameter: 3.2 mm (1.8") length: 38mm (1.5") |
XX denotes length of probe in feet.
3 m(10') and 7.6 m (25') probes available. |
2mm (0.07") Diameter Teflon coated cable |
V Range external probe:EPT-22W-XXV |
Sealed Teflon Tip Diameter: 3mm (0.12") Length: 28mm (1.1") |
Temperature Probe Accessories
Thermal Dampening Block for use in refrigerators and freezers. Simulates a glycol bottle to reduce viewLinc alarms generated by opening and closing a door.